SC-16890 General Ventilation Air Filter Test System
Application
It is used for the performance test of air filters for general ventilation, including resistance to airflow test, fractional efficiency test and dust capacity test. Test indicators: initial pressure differential, pressure differential to air flow rate, initial fractional efficiency, conditioned fractional efficiency, average fractional efficiency, pressure differential as a function of the test dust captured, arrestance as a function of the test dust captured.
Technical Parameters
Technical Data Sheet |
Item |
Parameter |
Output air volume |
500-5500m3 / h |
Maximum size of the filter can be tested |
6100*610mm |
Aerosol |
DEHS or KCL |
Filter efficiency range |
30%~90% |
Sampling flow |
≥2.83L/min |
Resistance range |
0~1000Pa |
Resistance accuracy |
±1% |
Dimension |
13700*2200*2400mm |
Product Pictures
Features
Designed and manufactured according to ISO 16890.
Adopt imported particle counters, differential pressure meter and other components to ensure the test accuracy.
Equipped with independent IPA discharge cabinet.
Test report printing, label printing, and database management functions are available.
1. Are you a trading company or manufacturer?
we are a professional tester manufacturer with 25 years of industry experience. Our factory is located in Suzhou, where is only 100 kilometers far away from Shanghai Hongqiao International Airport. It is pretty convenient for you to make a visit.
2. What is the L/T?
The lead time will be between15 and 30 days after deposit payment. If there is stock for the instrument you want, timely shipment will be arranged within 7 days.
3. What is your terms of payment?
Based on our sales policy, T/T 30% as deposit should be paid before order, and 70% balance payment should be completed before shipment. We will provide you with real photos or video of the instrument before balance payment.
4. What is your terms of delivery?
Available in EXW, FOB, CFR and CIF.
5. What is your packing for the tester?
Generally, packed with wooden case. We can do some adjustments in terms of your special requirement.
6. Can the mask tester be certified?
There is certification requirements for mask only, and there has no relevant certification requirements for mask tester, but a calibration report is required. we will provide a calibration report with a randomizer for our mask test instrument. The calibration report is valid for one year. After then, we can also support the corresponding calibration service for small fee.
7. Do you have after-sales service?
We can provide on-site installation and commissioning service for you, but you need to bear the travel expense of our engineer. The warranty period will be one year. Free replacement parts will be supplied during this period. Only accessories and corresponding shipping charges will be charged out of the warranty period.
8. Is there an English interface for your instrument?
Our instrument can be customized on the base of your demand. For example, both mask tester and filter material tester can be displayed in Chinese, English and Korean interfaces. The instruction manual is available in both Chinese and English.